Title: | Effect of x-ray energy on the radiological image quality in propagation-based phase-contrast computed tomography of the breast |
Contributor(s): | Wan, Sarina (author); Arhatari, Benedicta D (author) ; Nesterets, Yakov (author); Mayo, Sheridan C (author); Thompson, Darren (author) ; Fox, Jane (author); Kumar, Beena (author); Prodanovic, Zdenka (author); Hausermann, Daniel (author); Maksimenko, Anton S (author); Hall, Christopher J (author); Dimmock, Matthew R (author); Pavlov, Konstantin M (author) ; Lockie, Darren (author); Rickard, Mary (author); Gadomkar, Ziba (author); Alaleh, Alaleh (author); Vafa, Elham (author); Peele, Andrew G (author); Quiney, Harry M (author); Lewis, Sarah J (author); Gureyev, Timur E (author) ; Brennan, Patrick C (author); Tavakoli Taba, Seyedamir (author) |
Publication Date: | 2021-07-12 |
DOI: | 10.1117/1.JMI.8.5.052108 |
Handle Link: | https://hdl.handle.net/1959.11/58382 |
Abstract: | | Purpose: Breast cancer is the most common cancer in women in developing and developed countries and is responsible for 15% of women’s cancer deaths worldwide. Conventional absorption-based breast imaging techniques lack sufficient contrast for comprehensive diagnosis. Propagation-based phase-contrast computed tomography (PB-CT) is a developing technique that exploits a more contrast-sensitive property of x-rays: x-ray refraction. X-ray absorption, refraction, and contrast-to-noise in the corresponding images depend on the x-ray energy used, for the same/fixed radiation dose. The aim of this paper is to explore the relationship between x-ray energy and radiological image quality in PB-CT imaging.
Approach: Thirty-nine mastectomy samples were scanned at the imaging and medical beamline at the Australian Synchrotron. Samples were scanned at various x-ray energies of 26, 28, 30, 32, 34, and 60 keV using a Hamamatsu Flat Panel detector at the same object-to-detector distance of 6 m and mean glandular dose of 4 mGy. A total of 132 image sets were produced for analysis. Seven observers rated PB-CT images against absorption-based CT (AB-CT) images of the same samples on a five-point scale. A visual grading characteristics (VGC) study was used to determine the difference in image quality.
Results: PB-CT images produced at 28, 30, 32, and 34 keV x-ray energies demonstrated statistically significant higher image quality than reference AB-CT images. The optimum x-ray energy, 30 keV, displayed the largest area under the curve(AUCVGC) of 0.754 (p = 0.009). This was followed by 32 keV (AUCVGC = 0.731, p ≤ 0.001), 34 keV (AUCVGC = 0.723, p ≤ 0.001), and 28 keV (AUCVGC = 0.654, p = 0.015).
Conclusions: An optimum energy range (around 30 keV) in the PB-CT technique allows for higher image quality at a dose comparable to conventional mammographic techniques. This results in improved radiological image quality compared with conventional techniques, which may ultimately lead to higher diagnostic efficacy and a reduction in breast cancer mortalities.
Publication Type: | Journal Article |
Source of Publication: | Journal of Medical Imaging, v.8 (5) |
Publisher: | SPIE - International Society for Optical Engineering |
Place of Publication: | United States of America |
ISSN: | 2329-4310 2329-4302 |
Fields of Research (FoR) 2020: | 5103 Classical physics |
Peer Reviewed: | Yes |
HERDC Category Description: | C1 Refereed Article in a Scholarly Journal |
Appears in Collections: | Journal Article School of Science and Technology
|